Portable Mesh Nebulizer Online UN208
Mesh Nebulizer User Manual
Welcome to Your Mesh Nebulizer
This device adopts the latest microporous ultrasonic atomizing technology that sprays liquid medication into aerosol/vapor to inhale directly, achieves the purpose of painless, rapid and effective treatment. This device is used by both adults and children who suffer from the following conditions:
- Asthma
- Chronic Obstructive Pulmonary
- Disease(COPD)
- Emphysema
- Chronic Bronchitis
- other respiratory diseases with obstructed airflow
- Patients on and off ventilation or other positive pressure breathing assistance
CAUTION
- Please only use pure soluble liquid in this device, DO NOT use purified water, oil, milk, or thick liquid.The
atomization volume varies with the concentration of the liquid.
- Make sure to clean the mesh plate after each use to avoid blockage, DO NOT touch the mesh with your hand,
brushes or any hard objects.
- DO NOT submerge the mesh nebulizer in any liquid, make sure it fully dries if any liquid spilled on or inside the device
before using.
- DO NOT place the device on heated surface.
- DO NOT turn on the device without liquid in the liquid container.
Description of device and accessories
How to open the medicine cup module
How to replace the atomizer head
Function
1.Three working modes:High, Medium, Low. Choose the modes by pressing the power button.
2.Press the power button for 5 seconds to start automatic cleaning
3.The LED indicator light turns green and yellow alternately when the device is in automatic cleaning mode.
4.The device will automatically shut off after the use of 20 minutes.
5.The atomizer can be disassembled and replaced separately.
Before Use
It is imperative for hygiene reasons that the device and accessories are cleaned and disinfected before every use.
If the therapy requires that various different liquid be inhaled in succession, ensure that the medicine cup module is rinsed after every use.
1.Open the container lid, and fill the container with an isotonic saline, DO NOT overfill, The maximum recommended fill amount is 10ml, then close the container lid.
- Attach other parts like the mouthpiece and mask to the device based on your need.
Use mouthpiece:place the device to your month and close your lips firmly around the mouthpiece.
Use mask:place it over your nose and mouth.
3.Press on the power button to choose your needed working mode, note: each mode will take different time to atomize all the liquid, for example:
5ml medicine
High Mode : take about 15 minutes
Medium Mode: take about 20 minutes
Low Mode : take about 30 minutes
4.Press the power button to start the device.
5.The mesh nebulizer is on green light shows that it is working perfectly.
6.Press the power button again if the device turn off automatically after using 20 minutes.
- The atomizer can be removed and replaced, remove the atomizer by rotating it anticlockwise and install the atomizer by rotating it clockwise, as shown in the previous picture.
How to Clean and Maintenance
1.To clean accessories : remove the mouthpiece and any accessories from the device, wipe or soak with alcohol wipe.
2.To clean the nebulizer : add 6ml of clean water to the container and switch on the device to start
automatic cleaning mode, removing any residue from the mesh plate.
- If the outside of the device needs cleaning, wipe with dry towel.
4.Return the mesh plate to the device after full cleaning and store in a dry and cool place.
Reviews